UV-CUBE INTEGRATOR

UV-CUBE INTEGRATOR View larger

$2,576.00

New product

The UV-CUBE INTEGRATOR is a self-contained, high quality UV measuring instrument.

+ 3D measuring
+ UV-Intensity mW/cm2
+ UV-Dose mJ/cm2
+ Various UV-ranges available
+ Permanent or triggered measuring mode*
+ USB ComPort (option)
+ Graphic chart on computer (option)

It is designed to measure and display peak UV intensity and UV dosage in the UV curing process. In the standard version it is equipped with one UV sensor for the measuring of:

Full UV 230 410 nm

With this total UV band peak intensity and dose measuring, most of the measuring requirements of UV curing applications can be covered. Due to its UV sensor and the integrated microprocessor the UV-Cube Integrator can measure and display the peak UV-intensity of the total UV spectrum (mW/cm2). Additionally, this UV-Cube Integrator is calculating the UV-dosage (mJ/cm2) of the UV energy supplied during the time of exposure of one measuring cycle. The UV-dosage is calculated as the total Integral of UV-dosage over the full UV spectral bands. Optionally it is available with an extra sensor for measuring temperatures from 0 to 230 F / 0 to 110 C.

The sensor is on one side of the unit, while the display is on the opposite side. After completion of the measuring cycle the measuring results can be scrolled through on the built in 2 x 16 digit LCD display. A special AUTO-OFF feature that turns off the unit automatically after one minute serves as energy saving and extension of the battery service life. The UV-Cube Integrator is optionally available equipped with an USB ComPort and an evaluation software for downloading the data to a computer to show, edit and store a history of the measuring results of the entire measuring cycle as graphic charts (mW/cm2) and (mJ cm2)

TECHNICAL DATA
Spectral range: UV 230 410 nm (Standard)
Max. Power Input 0 to 5,000 mW/cm2
Display: LCD, 2x16 digits
Display range: 0 to 36,000 mJ/cm2
Measuring range: 0 to 2,000 mW/cm2
Sampling rate: 0.005 sec (200/sec)
Measuring period: 30 sec.
Readiness phase: 120 sec.
Power source: 7.2 V NiMH Accu-Pack, 70 mA, re-chargeable
Power consumption: 20 A
Battery service life: 2,000 measuring cycles
Dimensions: 2.4 x 2.4 x 2.4 (60 x 60 x 60 mm)
Weight: approx. 17.5 ounce (500 g)
Operating temperature: 0 to 113 F / 0 to 45 C
Heat protection: Heat shield
Base Accuracy: 5 %

While on the conveyer belt, the UV-Cube Integrator can withstand max. 230 F / 110 C for up to 10 seconds. The temperature of the housing should not exceed 113 F / 45 C. Because of uneven radiation distribution of the UV light source and different type of construction of the measuring devices by different manufacturers, different readings may appear under the same measurement conditions.

CALIBRATION
In order to keep its full function and precision it is recommended to have re-calibration done once per year. Re-calibration will also be necessary after change of battery. PTB traceable calibration acc. to DIN EN ISO / IEC 17025 with certificate

Datasheet_UVPS.gif
Download Datasheet:
UV-CUBE MICROPROCESSOR INTEGRATOR.pdf

SupportDoc_UVPS.gif
Download Support Document:
Measuring UV-Terms and Help.pdf

 

-->
1229 W. Cortland St. · Chicago, IL 60614-4805 · Phone: 1-800-621-1296 OR 773-248-0099 · Fax: 1-800-99FAXUV OR +1-773-880-6647· Email: info@uvps.com © COPYRIGHT 2019 U V Process Supply, Inc.