UV-1300 COLOR RADIOMETER AND DOSIMETER PLUS TEMPERATURE

UV-1300 COLOR RADIOMETER AND DOSIMETER PLUS TEMPERATURE View larger

New product

The UV-1300 COLOR RADI0METER AND DOSIMETER PLUS TEMPERATURE is a self-contained, high quality UV measuring instrument.

RDMTRSPLBB_M007422-423_WebPic1.jpg



+ UV-A intensity mW/cm² + UV-A dose mJ/cm²
+ UV-B intensity mW/cm² + UV-B dose mJ/cm²
+ UV-C intensity mW/cm² + UV-C dose mJ/cm²
+ *UV-V intensity mW/cm² + UV-V dose mJ/cm²
+ Full UV intensity mW/cm² + Full UV dose mJ/cm²
+ Big color graphic display
+ Numerical and graphical display
+ Permanent or „triggered“ measuring mode*

It is designed to measure and record UV intensity and UV dosage in the UV curing process. Measuring results are indicated both, graphically and numerically on a big colour display. It is equipped with three different UV sensors for the individual measuring of

UV-A 315 – 410 nm
UV-B 280 – 315 nm
UV-C 230 – 280 nm
UV-V 395 – 445 nm*
UV - 230 – 410 nm

With these three different UV-bands most of the measuring requirements of UV curing applications can be covered. Due to its three different UV sensors and the integrated microprocessor the UV-1300 Colour Radiometer + Dosimeter can measure, record and display the peak of the UV-intensity (mW/cm2) for each UV-band individually plus the peak of total UV energy. Additionally, this instrument is calculating the UV-dosage (mJ/cm2) of the UV energy supplied during the time of exposure of one measuring cycle. The UV-dosage is calculated for each UV-band (UV-A, UV-B and UV-C or UV-V) individually and as total Integral of UV-dosage over all three UV-bands. This allows to determine not only the total energy, but also how that energy is delivered, i.e., what intensity and dose at what UV-band. Optionally it is available with an extra sensor for measuring temperatures from 0 to 230° F / 0 to 110° C.

*This instrument features a selectable „triggered mode“, i.e. the 30 sec recording cycle starts within a 120 second readiness phase not before the incident UV-intensity exceeds 2 mW/cm²

The three sensors are on the back of the unit which also serves as a heat shield. After completion of the measuring cycle the measuring results are instantly displayed numerically and graphically auto-scaled on the built in 45 x 34 mm (1.75” x 1.3”) TFT colour display. A special AUTO-OFF feature that turns off the unit automatically after one minute serves as energy saving and extension of the battery service life.

The UV-1300 Color ComPort Radiometer + Dosimeter comes with a SD Memory card used to store data replaces the USB comport. The measuring data is stored directly to the inserted SD card and a circumstancial data transfer to a PC is not necesary after each measuring cycle. The evaluation software has been modified and is more user-friendly compared to earlier versions.

TECHNICAL DATA
Spectral ranges: UV-A 315 – 410 nm
UV-B 280 – 315 nm
UV-C 230 – 280 nm or (UV-V 395 – 445 nm)*
UV 230 – 410 nm
Max. Power Input 0 to 5,000 mW/cm2
Measuring range: 0 to 2,000 mW/cm2 or 2,000 to 5,000 mW/cm2
Sampling rate: 0.005 sec (200/sec)
Recording cycle: 30 sec.
Readiness phase: 120 sec.
Display range: 0 to 36,000 mJ/cm2
Display: TFT Colour Display, 45 x 34 mm (1.75” x 1.3”)
Power source: 2 x long life 3.6 V Lithium Battery
Power consumption: 20 µA
Battery service life: 2,000 measuring cycles
Dimensions: Æ 5.5” (140 mm), height ½” (13 mm)
Weight: approx. 17.5 ounce (500 g)
Operating temperature: 32° to 113° F / 0 to 45° Celsius
Heat protection: Heat shield on back plate
Base Accuracy: ± 5 %
Equipped with re-chargable batteries and charging unit
SD memory card slot

While on the conveyer belt, the UV-1300 Colour Radiometer + Dosimeter can withstand max. 230° F / 110° C for up to 10 seconds. The temperature of the housing should not exceed 113° F / 45° C. Because of uneven radiation distribution of the UV light source and different type of construction of the measuring devices by different manufacturers, different readings may appear under the same measurement conditions.

CALIBRATION
In order to keep its full function and precision it is recommended to have re-calibration done once per year. Re-calibration will also be necessary after change of battery. PTB traceable calibration acc. to DIN EN ISO / IEC 17025 with certificate

Datasheet_UVPS.gif
Download Datasheet:
UV-1300 RADIOMETER AND DOSIMETER.PDF

SupportDoc_UVPS.gif
Download Support Document:
Measuring UV-Terms and Help.pdf

   

-->
1229 W. Cortland St. · Chicago, IL 60614-4805 · Phone: 1-800-621-1296 OR 773-248-0099 · Fax: 1-800-99FAXUV OR +1-773-880-6647· Email: info@uvps.com © COPYRIGHT 2019 U V Process Supply, Inc.