UV-PROBE INTEGRATOR MULTI

UV-PROBE INTEGRATOR MULTI View larger

$1,820.00

New product

The UV-PROBE INTERGATOR MULTI is a self-contained, high quality UV measuring instrument.

+ Up to 4 UV-sensors
+ Up to 4 different UV-bands
+ UV intensity mW/cm2
+ UV peak intensity mW/cm2
+ UV dose mJ/cm2
+ USB ComPort (Option)
+ Graphic chart on computer (Option)
+ Rechargeable batteries and battery charger included

It is designed to measure and display UV intensity in mW/cm2. An additional function is the scan of the peak value of UV-intensity in mW/cm2 and to measure the UV dose in mJ/cm2 within a pre-set period of 30/60 seconds. It is able to automatically detect up to four different UV-sensors in up to four different UV-bands (UV-A, UV-B, UV-C, UV-Diazo or full UV) Extra sensors (up to three) should be ordered right with the initial purchase in order to calibrate sensors to the base unit. Due to its UV sensor(s) and the integrated microprocessor the UV-Probe Integrator Multi can measure and display the peak UV-energy of up to four different UV bands (mW/cm2).

Additionally, this UV-Probe Integrator Multi is calculating the UV-dosage (mJ/cm2) of the UV energy supplied during the time of exposure of one measuring cycle. The UV-dosage is calculated as the total Integral of UV-dosage over the UV spectral band of the respective sensor connected. The removable, probe-type sensors are connected to the base unit by a cable of approx. 1 meter (40) length. In the function Direct the actual UV-intensity in mW/cm2 supplied to the sensor is measured. The function Scan will start a 30 second measuring cycle of both, UV-intensity and UV-dose. After completion of the measuring cycle the measuring results can be scrolled through on the built in 2 x 16 digit LCD display. A special AUTO-OFF feature that turns off the unit automatically after one minute serves as energy saving and extension of the battery service life.

All UV Probe Integrator Multis have re-chargeable batteries and come with a battery charger.

Optionally, this microprocessor integrator is available with an USB ComPort and an evaluation software for downloading the data to a computer to show, edit and store a history of the measuring results of the entire measuring cycle as graphic and numeric charts (mW/cm2) and (mJ/cm2).The UV-Probe Integrator Multi (ComPort) is available in five different measuring ranges:

UV-Probe Integrator Type 1 UV-V (Diazo) 350 460 nm
UV-Probe Integrator Type 2 UV-A 315 400 nm
UV-Probe Integrator Type 3 UV 230 410 nm (STD)
UV-Probe Integrator Type 4 UV-B 280 315 nm
UV-Probe Integrator Type 5 UV-C 230 280 nm

Extra sensors are avaliable as follows:
Sensor Type 1 UV-Diazo 350 460 nm
Sensor Type 2 UV-A 315 400 nm
Sensor Type 3 UV 230 410 nm (STD)
Sensor Type 4 UV-B 280 315 nm
Sensor Type 5 UV-C 230 280 nm

UV-ComPort Probe Integrator Multi
UV-ComPort Probe Integrator Multi, Type 1 UV-V (Diazo) 350 460 nm
UV-ComPort Probe Integrator Multi, Type 2 UV-A 315 400 nm
UV-ComPort Probe Integrator Multi, Type 3 UV 250 410 nm
UV-ComPort Probe Integrator Multi, Type 4 UV-B 280 315 nm
UV-ComPort Probe Integrator Multi, Type 5 UV-C 230 280 nm

TECHNICAL DATA
Spectral ranges: UV 230 410 nm (Standard) or other
Max. Power Input 0 to 5,000 mW/cm2
Display: LCD, 2x16 digits
Display range: 0 to 60,000 mJ/cm2
Measuring range: 0 to 2,000 mW/cm2
Sampling rate: 0.005 sec (200/sec)
Recording cycle: 30 sec.
Power source: 2 x long life 3.6 V Lithium Battery
Power consumption: 20 A
Battery service life: 2,000 hrs
Dimensions: 5.5 (120 mm) x 3 (75 mm) x 0.4(10 mm)
Weight: approx. 6 ounce (150 g)
Dimensions of probe: 1.5 (40 mm) x 0.4 (10 mm)
Length of probe cable: approx. 40 ( 1 meter )
Operating temperature: 32 to 113 F / 0 to 45 Centigrade
Base Accuracy: 5 %

While measuring, the sensor of the UV-Probe Integrator Multi can withstand max. 230 F / 110 Centigrade for up to 10 seconds. The temperature of the housing should not exceed 113 F / 45 Centigrade. Because of uneven radiation distribution of the UV light source and different type of construction of the measuring devices by different manufacturers, different readings may appear under the same measurement conditions.


CALIBRATION
In order to keep its full function and precision it is recommended to have re-calibration done once per year. Re-calibration will also be necessary after change of battery. PTB traceable calibration with certificate.